EU MDR Software Classification Rule and FDA Comparison
Under EU MDR 2017/745, what classification rule applies to standalone software intended to inform decisions for diagnosis, and what class results? Compare with the FDA approach.
Under the EU MDR, standalone software intended to provide information used to take diagnostic or therapeutic decisions falls under Rule 11 of Annex VIII. It starts at Class IIa. It moves up to Class IIb if those decisions could cause a serious deterioration of a person's health or a surgical intervention, and to Class III if they could cause death or an irreversible deterioration. The FDA works differently. It first asks whether the software function is a device at all; some clinical decision support (CDS) software is excluded by statute. Software that is a device then goes into the ordinary Class I/II/III system and the matching route to market.
Approach: I searched the EU MDR/IVDR implementing and delegated acts. They don't contain the base text of Regulation (EU) 2017/745, so the Rule 11 wording below comes from Commission MDCG guidance and the US position from fda.gov. All of it is from search-result snippets, not the full documents.
EU MDR: Rule 11 (Annex VIII)
| Software purpose / possible impact | Class |
|---|---|
| Provides information used to take decisions for diagnosis or therapy (default) | IIa |
| Those decisions could cause a serious deterioration of health or a surgical intervention | IIb |
| Those decisions could cause death or an irreversible deterioration of health | III |
- Rule 11 also covers software that monitors physiological processes, which is a separate branch of the rule.
- MDCG 2019-11 explains that Rule 11 sets the class by combining two things: how important the software's information is to the healthcare decision, and how serious the healthcare situation or the patient's condition is.
- Revision 1 of MDCG 2019-11 (June 2025) added clarifications on Rule 11.
- Notified bodies are designated for software under the code "MDA 0315 Software" in Implementing Regulation (EU) 2017/2185 [1]. This matters in practice because Class IIa and above need a notified body.
FDA: decide first whether it is a device, then classify
Step 1: Is it excluded from the device definition?
-
Section 520(o)(1)(E) of the FD&C Act excludes CDS software functions, but only if all four statutory criteria are met. FDA's CDS guidance applies them function by function. In outline, the software must not:
- acquire, process or analyse a medical image or a signal from an in vitro diagnostic device or a signal acquisition system;
- do anything beyond displaying, analysing or printing medical information.
It must also be meant to support a healthcare professional (HCP), and let the HCP independently review the basis for its recommendations so they don't rely primarily on them.
-
FDA's revised final guidance also sets out an enforcement-discretion policy. It covers functions that fail the third criterion only because they give a specific diagnostic or treatment output, where just one clinically appropriate recommendation exists and all the other criteria are met.
Step 2: If it is a device, it is classified like any other device:
- Class I: general controls.
- Class II: general controls plus special controls.
- Class III: Premarket Approval (PMA).
- Non-exempt Class I and Class II devices need a 510(k).
Key differences
| Point | EU MDR | FDA |
|---|---|---|
| Rule for software | One software-specific rule (Rule 11) | No software-specific rule; general device classification applies |
| Default for diagnosis-informing software | Class IIa minimum | Can be outside the device definition if it meets all four Non-Device CDS criteria |
| What drives a higher class | Severity of harm from the decision (serious → IIb; death or irreversible → III) | Level of control needed for safety and effectiveness (I/II/III) |
| Weight given to HCP review | Doesn't lower the class | Central: if the HCP can independently review the basis, the function can be excluded |
The upshot: a diagnostic-support tool that FDA might treat as non-device CDS would still normally be at least Class IIa under the MDR and need a notified body.